Faults and Trouble-shooting methods for ZAMT-80G Medical Ozone Therapy Unit for Gynecology
Faults and troubleshooting methods: If you still can't eliminate the faults by referring to the following methods, please contact after-sales service personnel from Qianyan Medical to arrange the equipment repair.
Faults | Reason | Methods |
equipment without power | 1.fusing | replace cartridge fuse |
2.socket no-voltage or power cord obstructed | replace socket or inspect circut | |
abnormality of the equipment | 1.low supply voltage | Assure 200V-240V of supply voltage |
2.strong electric or magnetic field interference | move the equipment away | |
unable to get through the pressure test | 1.oxygen supply pipe leakage or broken | inspect oxygen supply pipe and replace it |
2.improper adjustment of pressure reducer | shift the pressure reducer to specified position | |
3.oxygen underpressure | assure oxygen cylinder be turned on or replace it | |
ozone concentration no display, unstable or cannot reach the set value | 1.using central oxygen supply | changed to bottled medical grade oxygen |
2.improper adjustment of pressure reducer | shift the pressure reducer to specified position |