Faults and Trouble-shooting methods for ZAMT-80 Medical Ozone Therapy Unit
Faults and troubleshooting methods: If you still can't eliminate the faults by referring to the following methods, please contact after-sales service personnel from Qianyan Medical to arrange the equipment repair.
Faults | Reason | Methods |
equipment without power | 1、fusing | replace cartridge fuse |
2、socket no-voltage or power cord obstructed | replace socket or inspect circuit | |
abnormality of the equipment | 1、low supply voltage | Assure 200V-240V of supply voltage |
2.strong electric or magnetic field interference | move the equipment away | |
unable to get through the pressure test | 1、oxygen supply pipe leakage or broken | inspect oxygen supply pipe and replace it |
2、improper adjustment of pressure reducer | shift the pressure reducer to specified position | |
3、oxygen underpressure | assure oxygen cylinder be turned on or replace it | |
ozone concentration no display, unstable or cannot reach the set value | 1、using central oxygen supply | changed to bottled medical grade oxygen |
2、improper adjustment of pressure reducer | shift the pressure reducer to specified position | |
cannot take ozone | 1、oxygen supply pipe leakage | inspect oxygen supply pipe and replace it |
2、syringe port fault | Replace O-ring according to the instruction manual | |
ozone leakage | 1、leakage from syringe port | Replace O-ring according to the instruction manual |
gas nozzle too tight | 1、damage of inside O-ring | Replace O-ring according to the instruction manual |
2、long time no use or new equipment | press the motion bar in the nozzle repeatedly | |
3、spring not to get up | replace the spring |